i-STAT Blood Gas Analyzer (POCT) 2



i-STAT 1 Analyzer features
•      Operating temperature:       16 to 30°C
            i-STAT 1 Analyzer tidak akan dapat berfungsi bila temperatur melebihi 30°C.
•      Batteries:
            Two x 9 Volt lithium batteries
            Rechargeable batteries dapat digunakan (dengan ketentuan daya harus 9 Volt)
Administration Menu
Tekan tombol MENU untuk masuk ke administration menu
Terdiri dari pilihan :
•      1 - Analyzer Status
•      2 - Data Review
•      3 - Quality Tests
•      4 - Customisation
•      5 - Set Clock
•      6 - Transmit Data
•      7 – Utility
Administration Menu:
1 - Analyzer Status
•      Temp              Room temperature
•      Pressure                    Barometric pressure
•      Battery                        Battery voltage
•      Uses               Total number of test cycles
•      Serial              Serial number of the Analyzer
•      CLEW                        Software version of standardization data
•      Version                      Version of software installed
•      Custom                      Customization profile name
•      Stored Records        Number of test records in memory
•      Unsent                       Number not transmitted to the CDS
Administration Menu:
2 - Data Review
•      1 - Patient                  Recall by scanning or entering patient ID
•      2 - Control                 Recall of aqueous control data
•      3 - Proficiency          Recall of external aqueous control data
•      4 - Cal Ver                 Calibration Verification data (only for US)
•      5 - Simulator             Recall of internal / external simulator data
•      6 - All              All tests records in the analyzer’s memory
•      7 - List                        A list of records in analyzer (easiest to use)
Administration Menu:
 3 - Quality Tests
•      1 - Control                 Stores aqueous QC results
•      2 - Proficiency          Stores external QC results
•      3 - Cal Ver                 Calibration Verification data (US only)
•      4 - Simulator Stores internal & external simulator results
Administration Menu:
 4 – Customisation
Password protected.  Default setting is no password – press Enter
•      1 - View
•      2 - Change
Leads to small sub menu:
•      1 - Analyzer
•      2 - ID Entry
•      3 - Patient Tests
•      4 - QC Tests
•      5 - Results
Administration Menu:
5 - Set Clock
Needs to be changed only for “daylight savings time”
Password protected
Default setting is no password – just press Enter key
•      Move cursor under number to change
•      Insert correct number with keypad
•      Press Enter key
Administration Menu:
6 - Transmit Data
Recommend that machine is set to autotransmit in customisation
This option only needed for manual transmission of data
•      1 - Most Recent
•      2 - This Month
•      3 - Last Month
•      4 - All
Administration Menu:
 7 – Utility
Password protected.
Default setting is no password – just press Enter key
•      1 – Send software
            To send software from one Analyzer to another
•      2 – Clear memory
            Typically used to clear a demonstration analyzer of old data
•      3 – Receive software
            To accept software transmitted across a hospital network
Penyimpanan Cartridge
•      Simpan dalam suhu 2 and 8°C sampai tiba waktu kadaluarsa
•      Tidak diperbolehkan memasukkan kembali ke lemari pendingin setelah cartridge dikeluarkan
•      Untuk 1 buah cartridge dihangatkan selama 5 menit (sebelum dilakukan pemeriksaan)
•      Untuk 1 kotak cartridge dihangatkan selama 1 jam (sebelum dilakukan pemeriksaan)
•      Untuk jenis cartridge Non-Blood Gas simpan dalam suhu ruangan selama 14 hari (<30°C) sedangkan untuk jenis cartridge Blood Gas dapat disimpan selama 2 bulan
Bagaimana cara SENSOR mengukur?
•      3 types of sensors:
–     POTENTIOMETRIC
•      Measure the differences in potential as compared to reference
•      Example: ion selective electrode (results from Na+, K+, Cl-…)
–     AMPEROMETRIC
•      Measure current generated by oxidation and reduction reactions
•      Example: lactate results
–     CONDUCTOMETRIC
•      Current is applied across the test solution and the magnitude of voltage different is measured
•      Example: Hematocrit results
Cara Memegang Cartridge
•      Lepaskan cartridge dari kemasan tanpa menyentuh bagian atas dan bagian tengah dari cartridge.
•      Pegang cartridge di bagian sisi kanan dan kiri!
•      Tidak diperbolehkan memegang bagian sensor
Mencetak Hasil dengan Martel Printer
  1. Tekan tombol ON pada printer
  2. Sejajarkan IR Windows Analyzer  dengan IR Windows Printer
  3. Jarak optimum adalah 3 cm
  4. Tekan tombol PRINT pada Analyzer
  5. Jangan merubah posisi Analyzer & Printer hingga proses selesai