Saturday, 12 December 2015

i-STAT Blood Gas Analyzer (POCT) 2



i-STAT 1 Analyzer features
      Operating temperature:       16 to 30°C
            i-STAT 1 Analyzer tidak akan dapat berfungsi bila temperatur melebihi 30°C.
      Batteries:
            Two x 9 Volt lithium batteries
            Rechargeable batteries dapat digunakan (dengan ketentuan daya harus 9 Volt)
Administration Menu
Tekan tombol MENU untuk masuk ke administration menu
Terdiri dari pilihan :
      1 - Analyzer Status
      2 - Data Review
      3 - Quality Tests
      4 - Customisation
      5 - Set Clock
      6 - Transmit Data
      7 – Utility
Administration Menu:
1 - Analyzer Status
      Temp              Room temperature
      Pressure                    Barometric pressure
      Battery                        Battery voltage
      Uses               Total number of test cycles
      Serial              Serial number of the Analyzer
      CLEW                        Software version of standardization data
      Version                      Version of software installed
      Custom                      Customization profile name
      Stored Records        Number of test records in memory
      Unsent                       Number not transmitted to the CDS
Administration Menu:
2 - Data Review
      1 - Patient                  Recall by scanning or entering patient ID
      2 - Control                 Recall of aqueous control data
      3 - Proficiency          Recall of external aqueous control data
      4 - Cal Ver                 Calibration Verification data (only for US)
      5 - Simulator             Recall of internal / external simulator data
      6 - All              All tests records in the analyzer’s memory
      7 - List                        A list of records in analyzer (easiest to use)
Administration Menu:
 3 - Quality Tests
      1 - Control                 Stores aqueous QC results
      2 - Proficiency          Stores external QC results
      3 - Cal Ver                 Calibration Verification data (US only)
      4 - Simulator Stores internal & external simulator results
Administration Menu:
 4 – Customisation
Password protected.  Default setting is no password – press Enter
      1 - View
      2 - Change
Leads to small sub menu:
      1 - Analyzer
      2 - ID Entry
      3 - Patient Tests
      4 - QC Tests
      5 - Results
Administration Menu:
5 - Set Clock
Needs to be changed only for “daylight savings time”
Password protected
Default setting is no password – just press Enter key
      Move cursor under number to change
      Insert correct number with keypad
      Press Enter key
Administration Menu:
6 - Transmit Data
Recommend that machine is set to autotransmit in customisation
This option only needed for manual transmission of data
      1 - Most Recent
      2 - This Month
      3 - Last Month
      4 - All
Administration Menu:
 7 – Utility
Password protected.
Default setting is no password – just press Enter key
      1 – Send software
            To send software from one Analyzer to another
      2 – Clear memory
            Typically used to clear a demonstration analyzer of old data
      3 – Receive software
            To accept software transmitted across a hospital network
Penyimpanan Cartridge
      Simpan dalam suhu 2 and 8°C sampai tiba waktu kadaluarsa
      Tidak diperbolehkan memasukkan kembali ke lemari pendingin setelah cartridge dikeluarkan
      Untuk 1 buah cartridge dihangatkan selama 5 menit (sebelum dilakukan pemeriksaan)
      Untuk 1 kotak cartridge dihangatkan selama 1 jam (sebelum dilakukan pemeriksaan)
      Untuk jenis cartridge Non-Blood Gas simpan dalam suhu ruangan selama 14 hari (<30°C) sedangkan untuk jenis cartridge Blood Gas dapat disimpan selama 2 bulan
Bagaimana cara SENSOR mengukur?
      3 types of sensors:
     POTENTIOMETRIC
      Measure the differences in potential as compared to reference
      Example: ion selective electrode (results from Na+, K+, Cl-…)
     AMPEROMETRIC
      Measure current generated by oxidation and reduction reactions
      Example: lactate results
     CONDUCTOMETRIC
      Current is applied across the test solution and the magnitude of voltage different is measured
      Example: Hematocrit results
Cara Memegang Cartridge
      Lepaskan cartridge dari kemasan tanpa menyentuh bagian atas dan bagian tengah dari cartridge.
      Pegang cartridge di bagian sisi kanan dan kiri!
      Tidak diperbolehkan memegang bagian sensor
Mencetak Hasil dengan Martel Printer
  1. Tekan tombol ON pada printer
  2. Sejajarkan IR Windows Analyzer  dengan IR Windows Printer
  3. Jarak optimum adalah 3 cm
  4. Tekan tombol PRINT pada Analyzer
  5. Jangan merubah posisi Analyzer & Printer hingga proses selesai

No comments: